Troubleshooting
Analytical Cytometer Problems:
SYMPTOM: No Signal with Calibration Beads
Possible causes: Improper size or FSC settings, FSC threshold too high, poor pressure, sample tube cracked sample line clog, laser problems
- Make sure instrument settings for beads have been located and SET into cytometer
- Make sure the sheath tank is pressurized, pressure valve should be UP (Scan) or away from you (Calibur)
- Make sure there is no crack in the tube
- Perform prime or drain and fill again
- Run 20% bleach on high for 15 minutes, followed by fresh DI water
SYMPTOM: Noisy FSC, no visible peak for beads
Possible causes: Debris or gunk at the intercept point, FSC threshold too low, air bubbles
- Run 20% bleach for 15 minutes, followed by fresh DI water.
- Check the threshold setting under CYTOMETER menu and ensure the threshold is defined for FSC and is set at 52.
- Drain and Fill or Prime again
SYMPTOM: Bead FSC signal OK when signals for other parameters are way off
Possible causes: Laser power fault, compensation values still engaged
- Open the compensation box under cytometer menu and ensure all settings are at 0.0%
- Open the status box under the cytometer menu and write down the laser power and laser current values with your sample tube on and fluidics in RUN
SYMPTOM: Red diode won't calibrate for FL4 Stains (on Calibur)
Possible causes: Red diode fault, FSC signal too low, event count too low.
- Turn fluidics to medium
- view particles on E01 with increase amplitude
- Make new solution of 1 drop BD APC beads in 1 ml saline or FACSFLOW
- Use APC compensation control to calibrate- set FL4 voltage to last recorded setting.
SYMPTOM: Continuous Standby Fault or Cytometer "Not Ready"
Possible causes: Alligator clips on sheath tank have slipped off, improper pressure levels, No sheath fluid, Laser Power fault
- Reseat the clips around the two metal prongs of male connector.
- Ensure sheath tank is full and pressurized
- Make sure sensors are properly tightened
- Check status box under cytometer menu and record laser power and current values
Software Problems:
SYMPTOM: Counters won't count: Save the template and close it. Quit Cell Quest. Reopen your template which will also open Cell Quest again. Locate Acquisition Control under Windows and proceed with acquisition.
SYMPTOM: Sample ID box blacks out when you try to write in it: Click in the Patient ID box and then click in the Sample ID box
SYMPTOM: Sentinel EvE error: Repeat what you were trying to do.
SYMPTOM: Unexpected quits with type "x" error: Reopen Cell Quest template you were working on. These errors are file construction errors/ file memory allocation errors.
SYMPTOM: Anything involving GPI0: Unfortunately, you'll have to shut down both computer and Cytometer and wait for help the next day or issue a dispatch call. See below
IF YOU CANNOT FIX THE PROBLEM:
- If your samples can remain unfixed on ice at 4C overnight, store them for running the next day.
- If not, you can try fixing your samples in 0.5% paraformaldehyde (as long as there is no PI in your sample).
YOU WILL BE HELD RESPONSIBLE FOR BENT OR BROKEN PIECES OR ANY DAMAGE INCURRED.
