Aria cell sorters

Aria II GC, BRU

Name Aria II GC, BRU
Location Room 2-203
Restricted use FCF staff operated only
Lasers 488 nm Blue (100 mW)
633 nm Red (30 mW)
355 nm UV (20 mW)
# of colours 11 colours (6B+SSC/3R/2UV)
Biosafety cabinet Yes
Default configuration "New advanced configuration"

"New Advanced Configuration" (default)

488nm 100 mW Solid State Blue (Coherent)
LP BP Fluorochromes
Detector A 755LP 780/60 PE-CY7
Detector B 685LP 720/40 PerCP-Cy5, PE-CY5.5 or PE-CY5.5, PerCP-eF710
Detector C 635LP 660/20 PE-CY5, PerCP
Detector D 600LP 610/20 PE-Texas Red or PI, 7AAD
Detector E 550LP 575/26 PE
Detector F 505LP 530/30 Alexa488, CFSE, FITC, GFP
Detector G - 488/10 SSC
Detector H - - no detector
633 nm 30 mW HeNe Red (JDS Uniphase)
Detector A 755LP 780/60 APC-CY7, APC-Alexa750, APC-Alexa780, APC-eF780
Detector B 690LP 710/50 Alexa 700
Detector C - 660/20 APC, Alexa633, Alexa 647
355 nm 20 mW DPSS UV (Mode-Lock)
Detector A 600LP 630/30 Hoeschst Red, Indo-1 (Blue)
Detector B - 424/44 DAPI, Fixable Blue, Hoescht Blue
Detector C - - No detector
Note: For Baseline and Performance checks the following filters are used on the Violet laser
Detector A 505LP 630/30 Hoescht Red, Indo-1 (Blue)

Note: PI may also be detected in Detectors B or C off the Blue laser

The "New Advanced Configuration" is the default cytometer configuration and was designed for a specific set of commonly used fluorochromes in which all detectors are used simultaneously. Filters in this configuration may not be optimal for your fluorochrome combination. Additional filters are available.