Becton Dickinson (BD) analyzers

LSRII-SC, VBR

BD LSR II
Name LSRII-SC, VBR
Location PGCRL, Room 18-9430
Restricted use No
Lasers 405 nm Violet (50 mW)
488 nm Blue (20 mW)
633 nm Red (40 mW)
# of colours 14 colours (5V/6B+SSC/3R)
HTS unit Yes
Default configuration "New advanced configuration 20160108"

"New Advanced Configuration 20160108" (default)

LP BP Fluorochromes
407 nm. 25 mW, DPSS Violet (Coherent)
Detector A 685 LP 710/50 BV711, Qdot705
Detector B 635 LP 660/20 BV650, Qdot655
Detector C 600 LP 610/20 BV605, Qdot605
Detector D 505 LP 525/50 BV510, V500, Amcyan, Fixable Aqua
Detector E - 450/50 BV421, V450, Pacific Blue, Fixable Violet, Sytox Blue, eF450
488nm, 20 mW, Solid State Blue (Coherent)
Detector A 755 LP 780/60 PE-Cy7
Detector B 685 LP 720/40 PerCP-Cy5.5, PE-Cy5.5, PerCP-eF710
Detector C 635 LP 660/20 PE-Cy5, 7AAD, PI (660/20)
Detector D 600 LP 610/20 PE-Texas Red, PI (610/20), PE/Dazzle 594, PE-CF564
Detector E 550 LP 575/26 PE
Detector F 505 LP 530/30 Alexa Fluor 488, CFSE, FITC, GFP, BB515
Detector G - 488/10 SSC
Detector H - - No detector
633nm, 40 mW, HeNe Red (JDS Uniphase)
Detector A 755 LP 780/60 APC-Cy7, APC-Alexa Fluor 750, APC-Alexa Fluor 780, APC-eF780
Detector B 685 LP 710/50 Alexa Fluor 700
Detector C - 660/20 APC, Alexa Fluor 633, Alexa Fluor 647

Note: For GFP, a 515/20 filter is optimal

The "Advanced Configuration" is the default cytometer configuration and was designed for a specific set of commonly used fluorochromes in which all detectors are used simultaneously. Filters in this configuration may not be optimal for your fluorochrome combination. Additional filters are available.