Aria cell sorters

Aria III CFI, BYRV

Name Aria III CFI, BYRV
Location Room 2-203
Restricted use FCF staff operated only
Lasers 488 nm Blue (15 mW)
561 nm Yellow/Green (50 mW)
633 nm Red (15 mW)
405 nm Violet (10 mW)
# of colours 17 colours (6B+SSC/2Y/3R/6V)
Biosafety cabinet Yes
Default configuration "New advanced configuration"

"New Advanced Configuration" (default)

488nm 15 mW Solid State Blue (Coherent)
LP BP Fluorochromes
Detector A 755LP 780/60 PE-CY7
Detector B 685LP 720/40 PerCP-Cy5, PE-CY5.5, PerCP-eF710
Detector C 635LP 660/20 PE-CY5, PerCP
Detector D 600LP 610/20 PE-Texas Red, PI, 7AAD
Detector E 556LP 576/26 PE
Detector F 502LP 530/30 Alexa488, CFSE, FITC, GFP
Detector G - 488/10 SSC
Detector H - - no detector
561 nm 50 mW Yellow/Green DPSS (Coherent)
Detector A 600LP 610/20 mCherry
Detector B - 582/15 DsRed
Detector C - - no detector
633 nm 15 mW HeNe Red (JDS Uniphase)
Detector A 755LP 780/60 APC-CY7, APC-Alexa750, APC-Alexa780, APC-eF780
Detector B 685LP 730/45 Alexa700
Detector C - 660/20 APC, Alexa633, Alexa647
407 nm 10 mW DPSS Violet (Coherent)
Detector A 735LP 780/60 QDot 800
Detector B 685LP 710/50 QDot 700
Detector C 630LP 670/14 QDot 655
Detector D 600LP 610/20 QDot 605
Detector E 505LP 525/50 Amcyan, Indo-1 unbound
Detector F - 450/50 DAPI, Hoescht, Cascade Blue, Pacific Blue, Alexa405
Detector G - - No detector
Detector H - - No detector

Note: PI may also be detected in Detectors B or C off the Blue laser
Note: If the Yellow/Green laser is on, a 582/15 filter is required in Detector E for detecting PE (and for running CST performance checks).

The "New Advanced Configuration" is the default cytometer configuration and was designed for a specific set of commonly used fluorochromes in which all detectors are used simultaneously. Filters in this configuration may not be optimal for your fluorochrome combination. Additional filters are available.