Beckman Coulter cell sorters

MoFlo XDP, UVBR

MoFlo XDP
MoFlo XDP
MoFlo XDP
MoFlo XDP
Name MoFloXDP, (U/V)BR
Location PGCRL Room 18-9430
Restricted use FMCF staff operated only
Lasers 355 nm UV (50 mW)
405 nm Violet (30 nW)
488 nm Blue (200 mW)
642 nm Red (75 mW)
# of colours 11-14 colours (2UV or 5V/6B+SSC/3R)
Biosafety cabinet Yes
Default configuration "Advanced configuration"

"New Advanced Configuration" (default)

DSP BP/LP Fluorochromes
355nm, 50mW, UV
FL7 505 450/65 Hoescht-Blue
FL8 - 630/30 Hoescht-Red
405nm, 30mW, Violet
FL7 505 450/65 Pacific Blue
FL8 - 530/40 Amcyan
FL9 585 630/30 Qdot 585
FL10 625 660/20 Qdot 655
FL11 680 710/50 Qdot 705
488nm, 200mW, Blue
SSC Detector 505 488/20 SSC
FL1 - 530/40 FITC
FL2 555 LP 580/30 PE
FL3 605 630/30 PE-Texas Red
FL4 650 670/30 PE-Cy5
FL5 680 695/40 PerCP-Cy5.5
FL6 733 740 LP PE-Cy7
642nm, 75mW, Red
FL12 700 LP 785/62 APC-Cy7
FL13 - 670/30 APC
FL14 695 DLP 720/13 Alexa Fluor 700
The "Advanced Configuration" is the default cytometer configuration and was designed for a specific set of commonly used fluorochromes in which all detectors are used simultaneously. Filters in this configuration may not be optimal for your fluorochrome combination. Additional filters are available.